608A11  Photo Industrial monitoring Instrumentation

Model 608A11
Product Type: Vibration Sensor, Accelerometer
Low-cost, general purpose, industrial ceramic shear ICP® accel, 100 mV/g, 0.5 to 10k Hz, top exit, 10-ft integral cable
View Spec Sheet (PDF)
View photo and drawing .
PERFORMANCE ENGLISH SI
Sensitivity(± 15 %) 100 mV/g
10.2 mV/(m/s²) [2]
Measurement Range ± 50 g
± 490 m/s²
Frequency Range(± 3 dB) 30 to 600,000 cpm
0.5 to 10,000 Hz
Resonant Frequency 1320 kcpm
22 kHz [1]
Broadband Resolution(1 to 10,000 Hz) 350 µg
3434 µm/sec2 [1]
Non-Linearity ± 1 %
± 1 % [3]
Transverse Sensitivity ≤ 7 %
≤ 7 %
ENVIRONMENTAL
Overload Limit(Shock) 5000 g pk
49,050 m/s² pk
Temperature Range -65 to +250 °F
-54 to +121 °C
Temperature Response See Graph
See Graph [1]
Enclosure Rating IP68
IP68
ELECTRICAL
Settling Time(within 1% of bias) ≤ 2.0 sec
≤ 2.0 sec
Discharge Time Constant ≥ 0.3 sec
≥ 0.3 sec
Excitation Voltage 18 to 28 VDC
18 to 28 VDC
Constant Current Excitation 2 to 20 mA
2 to 20 mA
Output Impedance <150 ohm
<150 ohm
Output Bias Voltage 8 to 12 VDC
8 to 12 VDC
Spectral Noise(10 Hz) 8 µg/√Hz
78.5 (µm/sec2)/√Hz [1]
Spectral Noise(100 Hz) 5 µg/√Hz
49.1 (µm/sec2)/√Hz [1]
Spectral Noise(1 kHz) 4 µg/√Hz
39.2 (µm/sec2)/√Hz [1]
Electrical Isolation(Case) >108 ohm
>108 ohm
PHYSICAL
Size (Hex x Height) 9/16 in x 2.5 in
14 mm x 64 mm
Weight(with cable) 3.5 oz
99.3 gm
Mounting Stud
Stud
Mounting Thread 1/4-28 Female
1/4-28 Female [4]
Mounting Torque 2 to 5 ft-lb
2.7 to 6.8 N-m
Sensing Element Ceramic
Ceramic
Sensing Geometry Shear
Shear
Housing Material Stainless Steel
Stainless Steel
Sealing Molded
Molded
Electrical Connector Molded Integral Cable
Molded Integral Cable
Electrical Connection Position Top
Top
Cable Length 10 ft
3 m
Cable Type Polyurethane
Polyurethane [5]
SUPPLIED ACCESSORIES:
Model 081A40 Mounting Stud (1)
Model ICS-2 NIST-traceable single-axis single-point amplitude response calibration at 6000 cpm (100 Hz) (1)
OPTIONAL VERSIONS
EX- Hazardous Area Approval- contact factory for specific approvals
Hazardous Area Approval EEx nL IIC T4, -40°C≤Ta≤121°C, II 3 G
Hazardous Area Approval EEx nL IIC T4, -40°C≤Ta≤121°C, II 1 G
Hazardous Area Approval Cl I, Div I, Groups A, B, C, D; Cl II, Div I, Groups E, F, G; Cl III, Div I
Hazardous Area Approval Cl I, Div 2, Groups A, B, C, D; ExnL IIC T4, AExnA IIC T4
Hazardous Area Approval Exia IIC T4, AExia IIC, T4
LB- Low Bias Voltage
Output Bias Voltage 4.8 to 6.5 VDC
Excitation Voltage 12 to 28 VDC
Measurement Range ± 10 g
± 98 m/s²
M- Metric Mount
Supplied Accessory : Model M081A61 Mounting Stud 1/4-28 to M6 X 1 (1) replaces Model 081A40
TO- Temperature Output
Temperature Output Range +36 to +250 °F
+2 to +121 °C
Temperature Scale Factor 5.56 mV/°F + 32
+10 mV/°C
Electrical Connector Molded Integral Cable
Electrical Connections(Red) Acceleration Output
Electrical Connections(Black) Ground
Electrical Connections(White) Temperature Output

All specifications are at room temperature unless otherwise specified.
NOTES:
[1] Typical.
[2] Conversion Factor 1g = 9.81 m/s².
[3] Zero-based, least-squares, straight line method.
[4] 1/4-28 has no equivalent in S.I. units.
[5] Twisted shielded pair.
[6] See PCB Declaration of Conformance PS023 or PS060 for details.



[6]



608A11  Photo



Download Drawing 12013-B.pdf
Top